Water Conductivity, Temperature and Level Sensors (DQA410 / DQA420)

Built for aggressive water such as landfill leachate, mine drainage, coastal intrusion and treatment plants, the DQA410 and DQA420 use a titanium body and four-electrode cell to stay accurate across a very wide conductivity range. Both report temperature-compensated conductivity, specific conductivity, salinity, TDS and resistivity; the DQA420 adds a precision level channel and water density. A five-year internal battery and three output options make them easy to integrate into any logger.
  • Four-electrode conductivity cell with two control electrodes, 5-100,000 uS/cm
  • Derived salinity (0-42 PSU), TDS, resistivity and, on DQA420, water density
  • Temperature -5 to 50 C with 0.1 C accuracy
  • DQA420 level channel 0-11 m with 0.05% accuracy and under 0.1% FS stability
  • RS485 Modbus, SDI-12 or 4-20 mA (single parameter) outputs
  • Titanium body and shatterproof construction for corrosive and abrasive water
  • 8-36 V DC supply or 3.6 V battery with five-year life
  • Cables from 15 to 80 m; calibration kit SVCKA2002 available
Conductivity range5-100,000 uS/cm
Conductivity accuracy+/-0.5% + 1 uS/cm (under 80,000); +/-1% above
Conductivity resolution0.1 uS/cm
Temperature-5 to 50 C, +/-0.1 C, 0.01 C resolution
Level (DQA420)0-11 m, +/-0.05%, 0.005% FS resolution
Derived parametersSpecific conductivity, salinity, TDS, resistivity, density (DQA420)
OutputsRS485 Modbus, SDI-12, 4-20 mA
Power supply8-36 V DC (15 mA) or 3.6 V battery, 5 years
MaterialsTitanium body, Delrin cone, PVC cell
Weight0.5 kg
  • Landfill leachate and remediation sites
  • Mine water and industrial discharges
  • Coastal and estuary saline intrusion
  • Sewage and water treatment plants
  • Groundwater quality networks