UV-A and UV-B Radiometers (DPA817 / DPA822)

The DPA817 (UV-A) and DPA822 (UV-B) use high-sensitivity photodiodes behind thin-film interference filters, with a glass dome and Teflon diffuser for good cosine response at low sun angles. They are designed to WMO No. 8 guidance for solar UV monitoring and are calibrated outdoors against reference instruments traceable to the ARPA Ivrea laboratory. LSI data loggers can combine the readings into an estimated UV Index.
  • Separate UV-A (DPA817) and UV-B (DPA822) sensors with broadband spectral response
  • Interference-filter photodiode with dome and diffuser for cosine error within 8% below 60 deg
  • Outdoor calibration traceable to ARPA Ivrea (ISO/IEC 17025)
  • 4-20 mA analogue output, 10-30 V AC/DC supply, 0.7 W consumption
  • UV Index estimate computed in E-Log and Alpha-Log loggers
  • Anodised aluminium IP66 housing; pole mounting with DYA034 arm and DYA049 collar
  • Recalibration interval of two years
Spectral range315-400 nm (DPA817); 280-315 nm (DPA822)
Measuring range0-70 W/m2 (DPA817); 0-5 W/m2 (DPA822)
Expanded accuracy (daily, clear sky)18% (DPA817); 22% (DPA822)
Cosine error+/-8% below 60 deg incidence
Output4-20 mA
Power supply10-30 V AC/DC, 0.7 W
Max load300 ohm
HousingAnodised aluminium, IP66
EMCEN 61326-1:2013
CableSold separately (5-100 m)
  • UV Index and public health warning
  • Meteorological and atmospheric monitoring
  • Materials weathering and exposure testing
  • Solar and environmental research