PQ200 Ambient Air Particulate Sampler

  • Portable sampler for TSP, PM10, PM2.5, PM Coarse, and PM1 
  • EPA Reference Method for PM10 sampling (Designation No. RFPS-1298-125) 
  • EPA PM2.5 Reference Method sampler (Designation No. RFPS-0498-116) 
  • EPA PM Coarse approved sampler 
  • Only designated reference sampler capable of a 24 hour run on internal built-in 12 volt battery. Internal battery provides power during AC power outages meaning you never have sample interruptions. 
  • Ordered by EPA as the PEP Audit Reference Method 
  • Only designated Reference Method sampler with solar power option 
  • Optional solar panel can allow for indefinite operation without charging or connection to an AC power source 
  • Single channel design for maximum reliability and portability 
  • Friendly operator interface and easy data collection and downloading 
  • Includes built-in 12-volt DC battery power system charger (U.L. approved), allows operation for up to 30 hours 
  • Advanced flow control system 

Particle Size

PM10, PM2.5, PM1, TSP 

Flow Range

10-20 LPM 

US EPA Designation

EPA RFPS-1298-125 FRM for PM10, EPA RFPS-0498-116 FRM for PM2.5, EPA RFPS-1208-173 FRM for PM10-2.5 

Sampling Interval

24-Hour Sampling

Pump

Dual Diaphragm Vacuum Pump 

 

Automatic, multi-point flow calibration using the deltaCal calibrator 

Accuracy

±2% 

Operating Temperature

-30 to 50° C 

Ambient/Filter Temperature Sensor Resolution

0.1° C 

Temperature Sensors Accuracy

±0.16° C 

Barometric Pressure Sensor

545 to 800 mmHg 

Internal Battery

12V 18Ah 

Data Output

RS232 & USB Connection 

 

USB connectivity for easier data download via portable flash drive

 

 

Easily calibrate temperature/pressure through simple-to-navigate menu

 

EPA Federal Reference Method for both PM10 and PM2.5 for variety of application such as remote isolated areas