The Kanomax 3719-A is the GMP-oriented variant of Kanomax’s remote particle sensor, counting the 0.5 and 5.0 µm sizes required for Grade A/B environmental monitoring. It samples at 1.0 CFM via an external vacuum source and streams counts, alarms and error flags over Modbus RS-485, with a 4–20 mA output for PLC or BMS integration. The compact stainless steel housing mounts easily at filling lines, isolators and other critical locations.
- Counts at 0.5 and 5.0 µm to match EU GMP Annex 1 limits
- 1.0 CFM (28.3 L/min) sampling from an external vacuum system
- Modbus over RS-485 and 4–20 mA analogue output
- Reports particle counts, alarm status and error conditions continuously
- Works standalone or as part of a facility monitoring system
- Powered via 24 V DC network hub or local 9–28 V adapter
- ISO 21501-4 and JIS B9921 compliant with CE marking
- Stainless steel enclosure and 3-year warranty
| Size channels | 0.5 and 5.0 µm |
| Flow rate | 1.0 CFM (28.3 L/min) |
| Counting efficiency | 50% ±20% at smallest channel |
| Coincidence loss | <10% at 500,000 particles/ft³ |
| Zero count | <1 count per 5 minutes |
| Interfaces | RS-485 (Modbus RTU); 4–20 mA analogue output |
| Power | 24 V DC via network hub, or 9–28 V DC adapter |
| Vacuum source | External pump required |
| Operating environment | 10–30 °C, 20–85% RH non-condensing |
| Dimensions / weight | 133 × 95 × 80 mm, 1.0 kg |
- EU GMP Annex 1 Grade A/B continuous monitoring
- Aseptic filling lines and isolators
- Cleanroom facility monitoring systems
- Hospital surgical suites
- Aerospace and food industry controlled areas