Kanomax 2360 Air Quality Monitor (particles, CO₂, temperature, RH)

The Kanomax 2360 bridges cleanroom particle counting and indoor air quality assessment in one stainless-steel handheld. It counts six user-selectable size channels between 0.3 and 25 µm at 0.1 CFM, converts counts to an estimated mass concentration, and simultaneously records CO₂, temperature and relative humidity. A 4.3-inch colour touchscreen, 45,000-sample memory with 1,000 location IDs and Ethernet, USB, RS-485/RS-232 and optional Wi-Fi connectivity support both surveys and fixed monitoring.
  • Six simultaneous size channels, user-selectable from 0.3 to 10 µm, total particulate to 25 µm
  • Estimated mass concentration (PM) via density and refractive-index settings
  • Integrated CO₂ (0–5,000 ppm), temperature and relative humidity sensors
  • 4.3-inch colour touchscreen with icon-based navigation and sample annotation
  • Stores 45,000+ samples, 1,000 locations and 50 sampling profiles
  • Ethernet, USB, RS-485/RS-232 and optional 802.11 b/g Wi-Fi
  • Built-in report generation for ISO 14644-1, EU GMP Annex 1 and FS 209E
  • ISO 21501-4 and JIS B9921 compliant; stainless steel enclosure
Measured parametersParticles, CO₂, temperature, relative humidity
Size range / channels0.3–25 µm; 6 channels (0.3, 0.5, 1.0, 2.5, 5.0, 10.0 µm, variable binning)
Flow rate0.1 CFM (2.83 L/min)
Mass concentration range0.01–20,000 µg/m³
CO₂0–5,000 ppm, ±1%, 1 ppm resolution
Temperature / RH0–50 °C ±0.5 °C; 15–90% RH ±2%
Display4.3-inch WQVGA colour touchscreen
Data storageUp to 45,000 samples
CommunicationEthernet, USB, RS-485/RS-232, optional Wi-Fi
Dimensions / weight13.3 × 9.7 × 23.9 cm, 1.8 kg
  • Cleanroom monitoring and certification (ISO 14644, GMP)
  • Indoor air quality assessments in commercial buildings
  • Industrial hygiene and workplace exposure surveys
  • Pharmaceutical and semiconductor manufacturing areas
  • HVAC commissioning and filter performance checks