The Hemera TN Analyser measures the three inorganic nitrogen species that make up total nitrogen in wastewater and displays each one alongside the TN sum. Two methods are offered: the standard version reads nitrite and nitrate directly in the liquid by UV FTLS spectroscopy while ammonium is measured in the gas phase, and the full gas-phase version converts all species for the lowest detection limit. Both are immune to turbidity, colour and salinity and respond in under a minute.
- Reports NH₄⁺, NO₂⁻, NO₃⁻ and total nitrogen from one instrument
- Patented UV FTLS method with no interference from turbidity, TSS, colour or salinity
- Method 1 (liquid + gas-phase): 1 mg/L detection limit, ≤2 % FS accuracy
- Method 2 (full gas-phase): 0.1 mg/L detection limit, ≤1 % FS accuracy
- Response time of one minute or less
- Minimal reagent use and low maintenance
- Wall-mounted L800 enclosure or ATEX/IECEx explosion-proof version for Zone 1 or 2
- 4–20 mA, relay alarm and RS485 Modbus outputs
| Parameters | TN, NH₄⁺, NO₂⁻, NO₃⁻ |
| Measurement methods | UV FTLS spectroscopy (liquid + gas-phase) or full gas-phase UV FTLS |
| Example range | 0–100 mg/L (custom ranges on request) |
| Accuracy | ≤2 % FS (method 1); ≤1 % FS (method 2) |
| Repeatability | ±2 % FS (method 1); ±1 % FS (method 2) |
| Detection limit | 1 mg/L (method 1); 0.1 mg/L (method 2) |
| Response time | ≤1 min |
| Enclosures | Wall-mounted (safe area); explosion-proof ATEX/IECEx Zone 1 or 2 |
| Outputs (L800 platform) | 4–20 mA isolated, programmable relays, RS485 Modbus |
| Power (L800 platform) | 100–240 VAC 50/60 Hz or 24 VDC, <20 W typical |
- Nitrification / denitrification control in biological treatment
- Wastewater discharge compliance
- Industrial effluent monitoring
- Eutrophication control in receiving waters
- Protection of downstream equipment