LSI LASTEM’s air temperature sensor is a Class AA platinum resistance probe that needs no power and connects directly to a data logger over four wires, so long cable runs do not affect the reading. Fitted inside a naturally or fan-ventilated radiation screen it becomes a full meteorological air temperature sensor; converters are available to turn the Pt100 signal into 4-20 mA or Modbus RS485.
- Pt100 Class AA (1/3 DIN B) element with ±0.1 K accuracy at 0 °C
- 4-wire connection cancels cable resistance over long runs
- Passive sensor: no power supply required
- Measuring range -50 to +70 °C, operating range -40 to +80 °C
- Optional natural or forced-ventilation radiation screens for outdoor use
- Optional converters for 4-20 mA (DEA420.1) or RS485 Modbus-RTU (MDMMA1010.1) output
- Supplied with free wires or Mini-DIN connector for LSI data loggers
- Calibrated in LSI’s ISO/IEC 17025-accredited laboratory
| Models | DMA033A, DMA033.1, EST033 |
| Sensing element | Pt100 Class AA IEC 60751, 4-wire |
| Measuring range | -50 to +70 °C |
| Accuracy | ±0.1 K at 0 °C |
| Resolution (with LSI logger) | 0.01 °C |
| Response time T90 | 30 s (bare); 6 min with protective filter at 0.2 m/s |
| Output | Pt100 (EN 60751 table); 4-20 mA or RS485 Modbus via converter |
| Power consumption | None |
| Protection / operating temp. | IP54 / -40 to +80 °C |
| Dimensions / weight | L 168 mm, Ø 18 mm / 0.44 kg |
- Automatic weather stations
- Agro-meteorology
- Indoor thermal comfort studies
- Building and HVAC monitoring
- Environmental research