The Kanomax 3081 is a battery-powered microbial air sampler for environmental monitoring in cleanrooms, hospitals and food facilities. It impacts airborne organisms onto standard culture media at 100 L/min with ±2.5% flow accuracy, programmes sample volumes from 20 to 6,000 litres, and stores up to 10,000 results as CSV for export over USB or RS-485. A remote controller lets operators run the sampler from outside the clean zone, and password-protected roles keep settings and data secure.
- Stable 100 L/min flow with ±2.5% accuracy for ISO 14698-1 style monitoring
- Hygienic SUS316 stainless steel impaction head
- Accepts 90–100 mm Petri dishes and 55–84 mm contact plates
- 4.3-inch colour touchscreen showing sampling time and volume
- Single or repeat modes: 2–300 cycles, 10 s to 1 h intervals
- Password-protected administrator and general-user access
- 10,000-record CSV memory with USB and RS-485 interfaces
- Remote start/pause/stop controller and optional external printer
| Sampling method | Inertial impaction |
| Flow rate | 100 L/min, ±2.5% |
| Sampling volume | 20–6,000 L |
| Cycles / interval | 2–300 or continuous; 10 s to 1 h between cycles |
| Media | Petri dish 90–100 mm OD; contact plate 55–84 mm OD |
| Display | 4.3-inch colour touchscreen (English, Chinese, Japanese) |
| Communication / storage | USB, RS-485; 10,000 records (CSV) |
| Battery / power | Li-ion up to 7 h, 2–3 h recharge; or 100–240 V AC |
| Operating environment | 0–40 °C, 0–90% RH non-condensing |
| Dimensions / weight | 136 × 235 × 133 mm, approx. 3.0 kg |
- Routine viable air monitoring in cleanrooms
- Pharmaceutical, biotech and sterile manufacturing
- Hospital and healthcare infection control
- Food, beverage and cosmetics production