Net radiation is the difference between the solar and atmospheric radiation reaching the ground and the short-wave reflected and long-wave emitted radiation leaving it. LSI LASTEM’s two-surface net radiometers use thermopiles, giving fast response and low sensitivity to wind compared with black-body designs, and are built for permanent outdoor exposure. The DPA266 four-component instrument measures each flux separately for scientific energy-balance work.
- Upper and lower thermopile surfaces measure net radiation over 0.3-50 um
- PRRDA0100A with 10 m cable and calibration certificate; PRRDA0130 with IP65 connector
- Microvolt or 4-20 mA output; RS485 Modbus available as an accessory
- 5% daily accuracy, stainless-steel IP66 housing, -40 to 80 C
- DPA266 four-component version with two pyranometers, two pyrgeometers and Pt100 compensation
- DPA266 internal 1.6 W heater prevents condensation on domes
- Pole mounting via DYA031 or MAPSA2000 arms and DYA049 collar
- Compatible with E-Log and Alpha-Log (via ALIEM) loggers
| Spectral range | 0.3-50 um |
| Principle | Thermopile |
| Range | -1500 to 1500 W/m2 (PRRDA0100A); -150 to 1500 W/m2 (PRRDA0130) |
| Accuracy | 5% daily |
| Output | uV or 4-20 mA; RS485 Modbus optional |
| Power supply | 10-30 V AC/DC, under 0.4 W |
| Housing | Stainless steel, IP66 |
| Operating temperature | -40 to 80 C |
| DPA266 short-wave | 0-2000 W/m2, 305-2800 nm, ISO 9060 Class C |
| DPA266 long-wave | 4.5-50 um pyrgeometers, 12 V DC 1.6 W heater |
- Surface energy-balance and evapotranspiration studies
- Agro-meteorology and irrigation research
- Climate and micro-meteorological research
- Urban heat-island monitoring