The DPA817 (UV-A) and DPA822 (UV-B) use high-sensitivity photodiodes behind thin-film interference filters, with a glass dome and Teflon diffuser for good cosine response at low sun angles. They are designed to WMO No. 8 guidance for solar UV monitoring and are calibrated outdoors against reference instruments traceable to the ARPA Ivrea laboratory. LSI data loggers can combine the readings into an estimated UV Index.
- Separate UV-A (DPA817) and UV-B (DPA822) sensors with broadband spectral response
- Interference-filter photodiode with dome and diffuser for cosine error within 8% below 60 deg
- Outdoor calibration traceable to ARPA Ivrea (ISO/IEC 17025)
- 4-20 mA analogue output, 10-30 V AC/DC supply, 0.7 W consumption
- UV Index estimate computed in E-Log and Alpha-Log loggers
- Anodised aluminium IP66 housing; pole mounting with DYA034 arm and DYA049 collar
- Recalibration interval of two years
| Spectral range | 315-400 nm (DPA817); 280-315 nm (DPA822) |
| Measuring range | 0-70 W/m2 (DPA817); 0-5 W/m2 (DPA822) |
| Expanded accuracy (daily, clear sky) | 18% (DPA817); 22% (DPA822) |
| Cosine error | +/-8% below 60 deg incidence |
| Output | 4-20 mA |
| Power supply | 10-30 V AC/DC, 0.7 W |
| Max load | 300 ohm |
| Housing | Anodised aluminium, IP66 |
| EMC | EN 61326-1:2013 |
| Cable | Sold separately (5-100 m) |
- UV Index and public health warning
- Meteorological and atmospheric monitoring
- Materials weathering and exposure testing
- Solar and environmental research