The PRPMA3100 uses light scattering to measure three particle size fractions at once from an air sample drawn in at 2 l/min by its integrated fan and tempered before measurement. An electrostatic precipitator with built-in high-voltage module allows periodic zero and reference checks, giving stable long-term readings. Its RS485 Modbus-RTU output and 5-35 V DC supply make it easy to integrate into an LSI Alpha-Log weather station or third-party system.
- Simultaneous PM1, PM2.5 and PM10 measurement
- Light-scattering optical principle with tempered air sample
- Electrostatic precipitator for automatic zero and reference correction
- Integrated fan sampling at 2 l/min
- Sampling interval adjustable from 5 minutes to 24 hours
- RS485 Modbus-RTU output, 5-35 V DC supply
- Operates -20 to +60 °C and 0-99 % RH
- Compact IP65 enclosure for outdoor weather station use
| Model | PRPMA3100 |
| Measured fractions | PM1, PM2.5, PM10 |
| Measuring range | 0 to 1000 µg/m³ |
| Accuracy PM1 / PM2.5 | ±5 µg/m³ + 5 % (0-100 µg/m³); ±10 % (100-1000 µg/m³) |
| Accuracy PM10 | ±25 µg/m³ (0-100 µg/m³); ±25 % (100-1000 µg/m³) |
| Sample flow | 2 l/min (integrated fan) |
| Sampling interval | 5 min to 24 h |
| Output | RS485 Modbus-RTU |
| Power supply | 5 to 35 V DC |
| Operating conditions / dimensions / weight | -20 to +60 °C, 0-99 % RH / 81 x 45 x 148 mm / 0.4 kg, IP65 |
- Ambient air quality monitoring
- Weather stations with air quality parameters
- Workshops, factories and production areas
- Office and indoor air quality
- Museums and archaeological sites