Hemera NH₃ (ammonia) Analyser

The Hemera Ammonium Analyser takes a different route from conventional ISE and colorimetric instruments: the sample is treated so that ammonium is released as ammonia gas, which is then read optically by UV FTLS spectroscopy. Because the measurement happens in the gas phase, solids, colour, salinity, pH swings and BTEX have no effect on the result. Reagent use is under 1 mL per reading and maintenance is needed only every six months, making it well suited to aeration control and discharge monitoring.
  • Gas-phase UV FTLS measurement with zero interference from turbidity, TSS, colour, salinity, pH or BTEX
  • Detection limit of 0.1 mg/L with ≤1 % FS accuracy on the example 0–100 mg/L range
  • Response time under one minute for real-time aeration control
  • Reagent consumption below 1 mL per measurement
  • Maintenance interval of six months
  • Wall-mounted L800 enclosure or ATEX/IECEx explosion-proof version for Zone 1 or 2
  • 4–20 mA, relay alarm and RS485 Modbus outputs
  • Can be combined with NO₂⁻/NO₃⁻ channels for total nitrogen
ParameterNH₃
Measurement methodUV FTLS spectroscopy
Example range0–100 ppm (custom ranges on request)
Accuracy≤1 % FS
Repeatability±1 % FS
Detection limit0.5 ppm
Response time≤10 s
EnclosuresWall-mounted or 19″ rack (safe area); explosion-proof ATEX/IECEx Zone 1 or 2
Outputs (G800 platform)4–20 mA isolated, programmable relays, RS485 Modbus
Power (G800 platform)100–240 VAC 50/60 Hz or 24 VDC, <20 W typical
  • Ammonia slip monitoring on SCR / SNCR de-NOx systems
  • Stack emissions from industrial and combustion plant
  • Ambient air quality monitoring
  • Process gas monitoring in chemical and fertiliser plants
  • Regulatory compliance