Model 6050T Trace Level CO Analyser

  • Large colour touch screen, front panel keypad, external keyboard and mouse
  • Menu driven software
  • Ethernet, USB and RS-232/485 ports
  • Front panel USB connections for peripheral devices and firmware updates
  • Four independent analog outputs with flexible ranges
  • 8 digital input/outputs (I/Os)
  • Remote access
  • Automatic temperature and pressure compensation
  • Comprehensive internal data logging
  • Internal memory for more than 2 years of data storage
  • Modbus RTU and TCP protocol
  • External Filter (back panel)
  • Calibration by dilution
  • Universal Line voltage in the range of 90-260 VAC, 50/60Hz
 

Ranges:  0–50 ppm User set

Noise:  < 0.02 ppm

Lower Detectable Limit:  < 0.04 ppm

Zero Drift:  < 0.1 ppm per 24 hours

Span Drift:  < 0.5% of reading/24 hr

Cycle Time:  1 sample/second

Precision:  <0.5 % of URL

Linearity:  <1 % of URL

Sample Flow Rate:  0.5 to 1.0 LPM

Certification:  US EPA: RFCA-0817-248

Communications:  Ethernet, USB and RS-232/485 Ports, Modbus protocol

  • Air Quality Monitoring (In/Out door)
  • R&D
  • Boiler Room
  • Pharmaceutical